Design and Wear Testing of a Temporomandibular Joint Prosthesis Articulation
J.-P. Van Loon
'TMJ Research Group, Department of Oral & Maxillofacial Surgery, Groningen University Hospital, Groningen, The Netherlands
G.J. Verkerke
Department of Biomedical Engineering, University of Groningen, Groningen, The Netherlands
M.P. De Vries
Department of Biomedical Engineering, University of Groningen, Groningen, The Netherlands
L.G.M. De Bont
TMJ Research Group, Department of Oral & Maxillofacial Surgery, Groningen University Hospital, Groningen, The Netherlands
As part of the development of a total temporomandibular joint prosthesis, a prosthesis articulation was designed. The articulation consists of a spherical head (ball) of the mandibular part, rotating against an enveloping ultra-high-molecular-weight polyethylene (UHMWPE) disc with a flat cranial side, which slides along the opposing skull part. The aim of this study was to determine the in vitro wear rate of the articulation, and to predict the in vivo wear rate from the results. Based on a disc thickness of 5 mm and a ball diameter of 8 mm, the stresses within the disc were calculated by means of a finite element computer model. The wear rate of the ball-disc articulation was determined by in vitro wear tests, with a stainless-steel ball rotating against a UHMWPE disc in a serum-based lubricant. Eight discs were tested for seven million cycles each. The in vitro wear rate of the disc-skull part articulation was calculated from the test results of the ball-disc articulation. The maximum Von Mises' stress was less than the yield strength of UHMWPE and, therefore, was sufficiently low. The in vitro wear rate of the ball-disc articulation was 0.47 mm3 per million cycles. The in vivo expected total wear rate is 0.65 mm3 per year, corresponding with a yearly decrease of disc thickness of 0.0094 mm. Although it is difficult to judge whether this wear rate is sufficiently low, because the influence of UHMWPE wear particles in the TMJ region is unknown, both the expected wear rate and the decrease of thickness appear to be acceptable.
Key Words: temporomandibular joint temporomandibular joint prosthesis wear testing prosthesis design UHMWPE.
REFERENCES
- Amstutz HC, Campbell P., Kossovsky N., Clarke IC (1992). Mechanism and clinical significance of wear debris-induced osteolysis. Clin Orthop 276:7-18.[Medline]
[Order article via Infotrieve]
- Atkinson JR, Dowson D., Isaac GH, Wroblewski BM (1985). Laboratory wear tests and clinical observations of the penetration of femoral heads into acetabular cups in total replacement hip joints. III: The measurement of internal volume changes in explanted Charnley sockets after 2-16 years in-vivo and the determination of wear factors. Wear 104:225-244.[CrossRef]
- Bartel DL, Bicknell VL, Wright TM (1986). The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components for total joint replacement. J Bone Jt Surg Am 68:1041-1051.[Abstract/Free Full Text]
- Bartel DL, Rawlinson JJ, Burstein AH, Ranawat CS, Flynn WF Jr (1995). Stresses in polyethylene components of contemporary total knee replacements. Clin Orthop 317:76-82.[Medline]
[Order article via Infotrieve]
- Bragdon CR, O'Conner DO, Lowenstein JD, Jasty M., Syniuta WD (1996). The importance of multidirectional motion on the wear of polyethylene. Proc Inst Mech Eng 210(Pt H):157-165.
- Clarke IC, Gustafson A., Jung H., Fujisawa A. (1996). Hip-simulator ranking of polyethylene wear: comparisons between ceramic heads of different sizes. Acta Orthop Scand 67:128-132.[Medline]
[Order article via Infotrieve]
- Fontenot MG, Kent JN ( 1992). In vitro wear performance of Proplast TMJ disc implants. J Oral Maxillofac Surg 50:133-139.[Medline]
[Order article via Infotrieve]
- Fornasier V., Wright J., Seligman J. (1991). The histomorphologic and morphometric study of asymptomatic hip arthroplasty. A postmortem stufy. Clin Orthop 271:272-282.
- Gibbs CH, Lundeen HC (1982). Jaw movements and forces during chewing and swallowing and their clinical significance. In: Advances in occlusion. Lundeen HC, Gibbs CH, editors. Boston: John Wright PSG Inc, pp. 2-32.
- Good VD, Clarke IC, Anissian L. (1996). Water and bovine serum lubrication compared in simulator PTFE/CoCr wear model. J Biomed Mater Res 33:275-283.[Medline]
[Order article via Infotrieve]
- Hirakawa K., Bauer TW, Stulberg BN, Wilde AH, Borden LS (1996). Characterization of debris adjacent to failed knee implants of 3 different designs. Clin Orthop 331:151-158.[CrossRef][Medline]
[Order article via Infotrieve]
- Howie DW, Haynes DR, Rogers SD, McGee MA, Pearcy MJ (1993). The response to particulate debris. Orthop Clin North Am 24:571-581.[Medline]
[Order article via Infotrieve]
- Isaac GH, Wroblewski BM, Atkinson JR, Dowson D. (1992). A tribological study of retrieved hip prostheses. Clin Orthop 276:115-125.[Medline]
[Order article via Infotrieve]
- Iso TR 9325 (1989). Implants for surgery-Partial and total hip joint prostheses-Recommendations for simulators for evaluation of hip joint prostheses. Report No.: ISO TR 9325. Geneve, Switzerland: International Organization for Standardization.
- Kabo JM, Gebhard JS, Loren G., Amstutz HC (1993). In vivo wear of polyethylene acetabular components. J Bone Jt Surg Am 75(B):254-258.
- Kent JN, Block MS, Halpern J., Fontenot MG (1993). Update on the Vitek partial and total temporomandibular joint systems. J Oral Maxillofac Surg 51:408-415.[Medline]
[Order article via Infotrieve]
- Kiehn CL, DesPrez JD, Converse CF (1979). Total prosthetic replacement of the temporomandibular joint. Ann Plast Surg 2:5-15.[Medline]
[Order article via Infotrieve]
- Kilgus DJ, Moreland JR, Fineman GA, Funahashi TT, Tipton JS (1991). Catastrophic wear of tibial polyethylene inserts. Clin Orthop 273:223-231.[Medline]
[Order article via Infotrieve]
- Kumar P., Oka M., Ikeuchi K., Shimizu K., Yamamuro T., Okumura H., et al. (1991). Low wear rate of UHMWPE against zirconia ceramic (Y-PSZ) in comparison to alumina ceramic and SUS 316L alloy. J Biomed Mater Res 25:813-828.[CrossRef][Medline]
[Order article via Infotrieve]
- Lancaster JG, Dowson D., Isaac GH, Fisher J. (1997). The wear of ultra-high molecular weight polyethylene sliding on metallic and ceramic counterfaces representative of current femoral surfaces in joint replacement. Proc Inst Mech Eng 211(Pt H): 17-24.
- Livermore J., Ilstrup D., Morrey B. (1990). Effect of femoral head size on wear of the polyethylene acetabular component. J Bone Jt Surg Am 72(A):518-528.[Abstract/Free Full Text]
- McKellop H., Clarke IC, Markolf KL, Amstutz HC (1978). Wear characteristics of UHMW polyethylene: a method for accurately measuring extremely low wear rates. J Biomed Mater Res 12:895-927.[Medline]
[Order article via Infotrieve]
- McKellop HA, Clarke IC, Markolf K., Amstutz H. (1981). Friction and wear properties of polymer, metal and ceramic prosthetic joint materials evaluated on a multichannel screening device. J Biomed Mater Res 15:619-653.[CrossRef][Medline]
[Order article via Infotrieve]
- Mjoberg B. (1994). Theories of wear and loosening in hip prostheses. Wear-induced loosening vs. loosening-induced wear-a review. Acta Orthop Scand 65:361-371.[Medline]
[Order article via Infotrieve]
- prEN 12563 (1996). Non active surgical implants-Joint replacement implants-Specific requirements for hip joint replacement implants. Report No.: prEN 12563. Brussels: European Committee for Standardization.
- prEN 12564 (1996). Non-active surgical implants-Joint replacement implants-Specific requirements for knee joint replacement implants. Report No.: prEN 12564. Brussels: European Committee for Standardization.
- Rose RM, Goldfarb HV, Ellis E., Crugnola AM (1983). On the pressure dependence of the wear of UHMWPE. Wear 92:99-111.[CrossRef]
- Rostlund T., Albrektsson B., Albrektsson T., McKellop H. (1989). Wear of ion-implanted pure titanium against UHMWPE. Biomaterials 10:176-181.
- Rostoker W., Galante JO (1979). Contact pressure dependence of wear rates of ultra high molecular weight polyethylene. J Biomed Mater Res 13:957-964.[CrossRef][Medline]
[Order article via Infotrieve]
- Schuller HM, Marti RK (1990). Ten-year socket wear in 66 hip arthroplasties. Ceramic versus metal heads. Acta Orthop Scand 61:240-243.[Medline]
[Order article via Infotrieve]
- Schulte KR, Callaghan JJ, Kelley SS, Johnston RC (1993). The outcome of Chamley total hip arthroplasty with cement after a minimum twenty-year follow-up. The results of one surgeon. J Bone Jt Surg Am 75(A):961-975.[Abstract/Free Full Text]
- Semlitsch M., Willert HG (1997). Clinical wear behaviour of ultra-high molecular weight polyethylene cups paired with metal and ceramic ball heads in comparison to metal-on-metal pairings of hip joint replacements. Proc Inst Mech Eng 211(Pt H):73-88.
- Sonnenburg I., Sonnenburg M. (1985). Total condylar prosthesis for alloplastic jaw articulation replacement. J Maxillofac Surg 13:131-135.[Medline]
[Order article via Infotrieve]
- van Loon JP, de Bont GM, Boering G. (1995). Evaluation of temporomandibular joint prostheses: review of the literature from 1946 to 1994 and implications for future prosthesis designs. J Oral Maxillofac Surg 53:984-996.[Medline]
[Order article via Infotrieve]
- van Loon JP, Otten E., Falkenstrom CH, de Bont LG, Verkerke GJ (1998). Loading of a unilateral temporomandibular joint prosthesis: a three-dimensional mathematical study. J Dent Res 77:1939-1947.
- van Loon JP, Falkenstrom CH, de Bont LG, Verkerke GJ, Stegenga B. (1999). The theoretical optimal center of rotation for a temporomandibular joint prosthesis: a three-dimensional kinematic study. J Dent Res 78:43-48.
- Wang A., Essner A., Stark C., Dumbleton JH (1996a). Comparison of the size and morphology of UHMWPE wear debris produced by a hip joint simulator under serum and water lubricated conditions. Biomaterials 17:865-871.
- Wang A., Stark C., Dumbleton JH (1996b). Mechanistic and morphological origins of ultra-high molecular weight polyethylene wear debris in total joint replacement prostheses. Proe Inst Mech Eng 210(Pt H):141-155.
- Weightman B., Light D. (1986). The effect of the surface finish of alumina and stainless steel on the wear rate of UHMW polyethylene. Biomaterials 7:20-24.
- Willert HG, Bertram H., Buchhorn GH (1990). Osteolysis in alloarthroplasty of the hip. The role of ultra-high molecular weight polyethylene wear particles. Clin Orthop 258:95-107.[Medline]
[Order article via Infotrieve]
- Wolford LM (1997). Temporomandibular joint devices: treatment factors and outcomes. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 83:143-149.[CrossRef][Medline]
[Order article via Infotrieve]
- Zichner LP, Willert HG (1992). Comparison of alumina-polyethylene and metal-polyethylene in clinical trials. Clin Orthop 282:86-94.[Medline]
[Order article via Infotrieve]
Journal of Dental Research, Vol. 79, No. 2,
715-721 (2000)
DOI: 10.1177/00220345000790020401

CiteULike Complore Connotea Del.icio.us Digg Reddit Technorati Twitter What's this?
|
|