Effects of Rare-earth Filters on Patient Exposure and Image ContrastDepartment of Dental Ecology, School of Dentistry, School of Dentistry, University of North Carolina, Chapel Hill, North Carolina 27514
Department of Radiology, School of Medicine, School of Dentistry, University of North Carolina, Chapel Hill, North Carolina 27514
Department of Oral Diagnosis, School of Dentistry, University of North Carolina, Chapel Hill, North Carolina 27514 Minimizing patient exposure while maintaining a diagnostically acceptable radiograph is a major goal in diagnostic radiography. Rare-earth filters may be the means to achieve this goal due to their "bandpass effect". The purpose of this study was to examine the image contrast effects and exposure reductions for various thicknesses of aluminum, samarium, gadolinium, gadolinium oxysulfide, and gadolinium oxysulfide added to 2.5 mm of aluminum. Trials were conducted on an intra-oral dental x-ray unit (range, 65 to 90 kVp). When compared with conventional aluminum, all of the rare-earth filters provided lower radiation exposures, with gadolinium in the metallic or oxysulfide form providing the lowest exposures. Samarium, at a thickness of 0.127 mm, yielded the highest image contrast. Gadolinium or gadolinium oxysulfide added to 2.5 mm of aluminum resulted in a slight loss of contrast when compared with conventional aluminum filtration. This loss may not be clinically significant, and when coupled with the reduced exposure afforded by these filters, they become viable as acceptable alternatives to aluminum filtration.
Journal of Dental Research, Vol. 66, No. 8,
1326-1330 (1987) |
|
|||